  CMS-II: Central Monitoring System
  Model TS4: Multiple Probe Temperature Monitor Alarm
  Model 140: Single Probe Temperature Monitor Alarm
  CA 5000: Freezer Monitor Alarm Backup System with LN2 or CO2 Injection

 

 

Hampshire Controls Corp. Product 

Laboratory Freezer Monitor Alarm Backup System with Liquid Nitrogen Injection, OR, Liquid Carbon Dioxide Injection.

Product Information: Model CA5000 and CA5000 N

Freezer Monitor Alarm System - Model CA 5000 by Hampshire Controls.Click to View Larger Image of CA5000, Freezer Monitor Alarm Backup System

Basic Features:

  • 24-hour battery back-up time in case of power failure.
  • Digital display of backup freezer or refrigerator temperature, temperature alarm set point, and current freezer operating temperature.
  • Relay for connecting to central monitoring system (CMS II), dialers or remote alarms.
  • Lid switch input for safe carbon dioxide (CO2) delivery (injection).
  • Large capacity, no maintenance battery with charger.
  • Exception value for functionality.

Laboratory Freezer Monitor Alarm Backup System with Either Liquid Carbon Dioxide Backup Injection OR Liquid Nitrogen Backup Injection Feature:

Models CA 5000 and CA 5000 N

Hampshire Controls Corporation's Model CA5000 Series Freezer Monitor Alarm Backup Systems are unique laboratory monitor and alarm devices that incorporate both field-proven, ultra-reliable freezer monitoring alarm capabilities with the unique ability to backup freezers in case of coolant leaks by injecting freezers with EITHER liquid carbon dioxide OR liquid nitrogen - in order to maintain laboratory freezer temperature. Either the liquid carbon dioxide or liquid nitrogen injection versions (model CA 5000 with CO2 injection and model CA 5000 N with liquid nitrogen injection) provide laboratories with an important additional level of safety when storing sensitive samples in their research freezers - since they "must" be maintained at specific ultra-low temperatures.

Model CA 5000 and CA S000 N Freezer Monitor Alarm Backup Systems offer user-programmable temperature alarm and control set-points (as well as the setting of different alarm delay times).  All operating parameters are stored in non-volatile memory, and thus are maintained even when power is lost.

The customer supplied liquid carbon dioxide or liquid nitrogen is maintained at high pressure; ready to be injected into the freezer should a malfunction occur (an alarm condition). This unique feature eliminates the warming of laboratory samples. Should a freezer coolant loss occur, the systems sense this, and a pre-determined amount of liquid carbon dioxide or liquid nitrogen is injected into the freezer to compensate for the loss of primary coolant - maintaining the freezers correct operation and protecting sensitive laboratory samples until the freezer is serviced.

The CA 5000 and CA 5000 N Freezer Alarm Monitor Backup Systems are simple to operate. Just plug the in, place the probe in the freezer to be monitored, and the display will show the temperature of that location.  Pushing three buttons allows the user to program various alarm set points as well as the additional protection. It's that simple - and THAT effective.

CA5000 Series Freezer Monitor Backup and Injection System Features:

* Internal battery with integrated charger - that keeps the system running for at least 24 hours after a power outage.
* "BAT" indicator that warns of a disconnected or low backup battery condition.
* Liquid Carbon Dioxide (CA 5000) version: Backup time is a minimum of 8 hours on one standard cylinder of CO2 at -60 °C set point.
* Liquid Nitrogen (CA5000 N) Version: Backup time of approximately 48 hours using one 65 liter cylinder of nitrogen at a -65 °C set point.
* User-programmable controls for freezer monitor alarm and for either injection system.
* Visible and audible indicators of system status.
* Multiple programmable monitor alarms - to minimize nuisance or false alarms.
* Relay outputs for central alarm systems or automatic dialers.
* A door safety interlock switch prevents gas injection while freezer is being accessed.

Freezer Monitor Alarm Backup and Injection Series System Documentation - Model CA5000 and CA 5000 N:

Freezer Monitor Alarm Backup and Injection Series - Model CA5000 and CA5000 N Pricing

Freezer Monitor Alarm Backup and Injection Series - Model CA5000 and CA5000 N Series Specifications - Consult Factory

Freezer Monitor Alarm Backup and Injection Series -Model CA5000 with Liquid Carbon Dioxide Injection Manual

Freezer Monitor Alarm Backup and Injection Series - Model CA5000 N with Liquid Nitrogen Injection Manual

All Documentation for Download

Hampshire Controls Corporation documentation has been setup to be read as a PDF using an Acrobat Reader plug-in on your browser. If necessary, please click the icon link below to install Adobe's Free Acrobat PDF Reader 9. For Firefox browser users, you first may need to make www.adobe.com an "allowed" URL.

Get Free Adobe PDF Reader

BACK TO MAIN PRODUCT LISTING PAGE


Hampshire Controls Corporation
1 Grove Street, PO Box 516, Dover, NH 03821-0516
(Call Toll Free in the US and Canada) 866-496-9424 - (Worldwide) 603-749-9424
(Fax) 603-749-9433
(Email) sales@hampshirecontrols.com

© Copyright, Hampshire Controls Corp..
na
na
na